Achieve precise molecular identification and quantification with the InfraLUM FT-08 FTIR Spectrometer, designed for routine mid-infrared analysis across diverse industries. This multipurpose Fourier transform instrument combines patented misalignment-proof optics with high-sensitivity detection—delivering exceptional signal-to-noise performance for chemical, pharmaceutical, petroleum and environmental applications. The hermetically sealed optical compartment with automated moisture monitoring ensures measurement stability, while the large sample area accommodates a wide range of accessories. Variable resolution from 0.5 to 64 cm⁻¹ handles everything from quick screening to detailed structural analysis. Built-in self-diagnostics and remote monitoring reduce downtime. SpectraLUM software with integrated calibration and library search modules simplifies operation for any skill level. Compatible with complete lines of Specac and PIKE Technologies accessories for transmission, reflection and ATR measurements. Share your analytical requirements—our team will configure the InfraLUM FT-08 with accessories and method libraries matched to your spectroscopy workflow.
Based on measuring infrared spectra using Fourier transform technique, the InfraLUM FT-08 employs a patented Double Cat’s Eye interferometer that is vibration-proof and misalignment-proof, ensuring long-term stability and reliable performance in routine laboratory environments without frequent realignment.
| Specification | Details |
| Spectral Range | 8,000–500 cm⁻¹ (ZnSe optics) |
| Variable Resolution | 64, 32, 16, 8, 4, 2, 1, 0.5 cm⁻¹ |
| RMS Signal-to-Noise Ratio | >60,000 (4 cm⁻¹, 1 min) |
| Stability of 100% Line | 0.2% |
| Scanning Speed | 0.8 s/scan at 16 cm⁻¹ resolution |
| Detector | High-resolution DLATGS |
| Beamsplitter | ZnSe in hermetically sealed desiccated housing |
| Wavenumber Accuracy | ±0.05 cm⁻¹ |
| Interferometer | Vibration- and misalignment-proof patented Double Cat’s Eye |
| Radiation Source | High-intensity long-life stabilized ceramic IR source |
| Sample Compartment | 230×240×180 mm |
| Power Consumption | 65 W |
| Weight | 32 kg |
| Dimensions (W×D×H) | 580×550×340 mm |
The objective is to cater to all verticals small & large scale industries by offering easy, simple & reliable instruments.