The Flash-170M is a manual Abel Closed Cup Flash Point Tester designed for accurate determination of the flash point of combustible liquids. Operating within a temperature range of -30°C to 70°C, it adheres to international standards such as ISO 13736 and IP 170. Ideal for laboratories requiring precise flash point determination of flammable liquids, the Flash-170M combines manual control with robust construction for reliable performance.

  • Manual Dipping Test: Allows for precise control over the testing process.
  • Heating System: Equipped with a 300W band heater operating at 230V AC for consistent heating.
  • Heat Rate Control: Features a precise ten-turn potentiometer for accurate heat rate adjustments.
  • Stirring Mechanism: Manual stirring with a speed control potentiometer, adjustable from 0 to 60 rpm.
  • Ignition Source: Gas igniter with precise control for pilot flame and ignition.
  • Cooling System: In-out ports for connecting to an external water or coolant circulator.
  • Sample Temperature Monitoring: Glass PT100 Class A thermometer with digital indication; thermometer is interchangeable.
  • Lid Assembly: Permanently fixed sample cup cover with stirrer, sample temperature sensor, and igniter for easy operation.
  • Compact Design: Dimensions: W: 220 mm x D: 300 mm x H: 325 mm; Weight: 17 kg.
  • Power Supply: 230V, 50Hz, 600W.
Specification Details
Model Flash-170M
Test Methods ISO 13736, IP 170
Temperature Range -30°C to 70°C
Heating System 300W band heater @ 230V AC
Heat Rate Control Ten-turn potentiometer
Stirring Mechanism Manual control, adjustable 0–60 rpm
Ignition Source Gas igniter with precise control
Cooling System In-out ports for external circulator
Sample Temperature Sensor Glass PT100 Class A with digital indication
Lid Assembly Fixed sample cup cover with stirrer, sensor, and igniter
Dimensions (W x D x H) 220 mm x 300 mm x 325 mm
Weight 17 kg
Power Supply 230V, 50Hz, 600W